The calibration process is dependent on the version of the scale, see date code information below.
To find your date code, refer to the product label on the back of the display head.
If still experiencing a problem after calibrating, please contact Health o Meter Technical Support at 800-638-3722 for troubleshooting assistance.

Calibration process for scales with a Date Code with the last two digits of 19 or lower:

Calibration of your scale is performed using kilograms (KG) or pounds (LB), according to the units of measure used upon entering into the calibration path. When calibrating in pounds, a total of 300 lbs is required, applied in increments of 100 lb, 200 lb and 300 lb. If calibrating in kilograms, a total of 120 kg is required, applied in increments of 40 kg, 80 kg and 120 kg.

Note: Only weights certified and traceable to national standards should be used for calibration procedures.

ACTIONDISPLAY SHOWS
Measurement Range 5 1/2 – 78 3/4 in (14 – 200 cm)
Graduation 1/8 inches (0.1 cm)